Evaluating the contact mechanics of human joints is an important element in understanding the pathomechanics of orthopaedic diseases. Although physical testing is essential in the evaluation process, reliable computational models… Click to show full abstract
Evaluating the contact mechanics of human joints is an important element in understanding the pathomechanics of orthopaedic diseases. Although physical testing is essential in the evaluation process, reliable computational models can augment these experiments by non-invasive predictions of biomechanical or surgical variables. The objective of this study was to perform verification of a framework for developing a medial forefoot finite element. Verification was conducted by comparing computational predictions to experimental measurements of first metatarsophalangeal and first metatarsocuneiform joint contact mechanics. A custom-built force-controlled cadaveric test-rig was used to derive measurements of contact pressure, force, and area. A quasi-static finite element was developed and driven under the same boundary and loading conditions. Calibration of cartilage moduli and mesh sensitivity analyses were performed. Mean errors in contact pressures, forces, and areas were 24%, 4%, and 40% at the first metatarsophalangeal joint and 23%, 12%, and 19% at the first Metatarsocuneiform joint, respectively. Verification of a medial forefoot finite element model development framework was presented and found to be within 30% for contact pressure and contact force of both joints. This study presents a method to verify and simulate realistic physiological loading to investigate orthopaedic diseases of the medial forefoot.
               