QAMaster: A new software framework for phantom‐based computed tomography quality assurance

Photo from wikipedia

Abstract The regular evaluation of imaging performance of computed tomography (CT) scanners is essential for CT quality assurance. For automation of this process, the software QAMaster was developed at the… Click to show full abstract

Abstract The regular evaluation of imaging performance of computed tomography (CT) scanners is essential for CT quality assurance. For automation of this process, the software QAMaster was developed at the Universitätsklinikum Erlangen, which provides based on CT scans of the CatPhan® 504 (The Phantom Laboratory, Salem, USA) automated image quality analysis and documentation by evaluating CT number accuracy, spatial linearity, uniformity, contrast‐noise‐ratio, spatial resolution, noise, and slice thickness. Dose assessment is supported by calculations of the weighted computed tomography dose index (CTDI w ) and weighted cone beam dose index (CBDI w ). QAMaster was tested with CatPhan® 504 scans and compared to manual evaluations of these scans, whereby high consistency of the respective results was observed. The CT numbers, spatial linearity, uniformity, contrast‐noise‐ratio, noise, and slice thickness deviated by only (0.13 ± 0.25) HU, (0.02 ± 0.05) mm, (−0.01 ± 0.03)%, 0.8 ± 1.8, (0.131 ± 0.05) HU, and (0.004 ± 0.005) mm between both evaluations, respectively. The QAMaster results for spatial resolution did not differ significantly (p = 0.34) from the CatPhan® 504 based manual resolution assessment. Dose computations were fully consistent between QAMaster and manual calculations. Thus, QAMaster proved to be a comprehensive and functional software for performing an automated CT quality assurance routine. QAMaster will be open‐source after its release.

Keywords: quality assurance; computed tomography; software; quality

Journal Title: Journal of Applied Clinical Medical Physics
Year Published: 2022

Link to full text (if available)


Share on Social Media:                               Sign Up to like & get
recommendations!

Related content

More Information              News              Social Media              Video              Recommended



                Click one of the above tabs to view related content.