Verification of patient-setup accuracy using a surface imaging system with steep measurement angle.

Photo from wikipedia

PURPOSE We evaluate an SGRT device (Voxelan HEV-600 M/RMS) installed with Radixact, with the view angle of the Voxelan's camera at 74 degrees. The accuracy of Voxelan with this steep angle… Click to show full abstract

PURPOSE We evaluate an SGRT device (Voxelan HEV-600 M/RMS) installed with Radixact, with the view angle of the Voxelan's camera at 74 degrees. The accuracy of Voxelan with this steep angle was evaluated with phantom experiments and inter-fractional setup errors of patients. METHODS In the phantom experiments, the difference between the measured values of Voxelan from the truth was evaluated for translations and rotations. The inter-fractional setup error between the setup using skin markers with laser localizer (laser setup: LS) and the setup using Voxelan (surface setup: SS) was compared for head and neck (N = 19), chest (N = 7) and pelvis (N = 9) cases. The inter-fractional setup error was calculated by subtracting from bone matching by megavoltage computed tomography (MVCT) as ground truth. RESULTS From the phantom experiments, the average difference between the measured values of Voxelan from the truth was within 1 mm and 1 degree. In all cases, inter-fractional setup error based on MVCT was not significantly different between LS and SS by Welch's t-test (P > 0.05). The vector offset of the LS for head and neck, chest, and pelvis were 6.5, 9.6, and 9.6 mm, respectively, and that of the SS were 5.8, 8.6, and 12.6 mm, respectively. Slight improvement was observed for the head and neck, and chest cases, however, pelvis cases were not improved because the umbilical region could not be clearly visualized as a reference. CONCLUSION The results show that SS in Voxelan with an installation angle of 74 degrees is equal to or better than LS.

Keywords: fractional setup; accuracy; surface; inter fractional; angle; setup

Journal Title: Journal of applied clinical medical physics
Year Published: 2022

Link to full text (if available)


Share on Social Media:                               Sign Up to like & get
recommendations!

Related content

More Information              News              Social Media              Video              Recommended



                Click one of the above tabs to view related content.