Thoracic motion‐compensated cone‐beam computed tomography in under 20 seconds on a fast‐rotating linac: A simulation study

Photo from wikipedia

Abstract Background Rapid kV cone‐beam computed tomography (CBCT) scans are achievable in under 20 s on select linear accelerator systems to generate volumetric images in three dimensions (3D). Daily pre‐treatment… Click to show full abstract

Abstract Background Rapid kV cone‐beam computed tomography (CBCT) scans are achievable in under 20 s on select linear accelerator systems to generate volumetric images in three dimensions (3D). Daily pre‐treatment four‐dimensional CBCT (4DCBCT) is recommended in image‐guided lung radiotherapy to mitigate the detrimental effects of respiratory motion on treatment quality. Purpose To demonstrate the potential for thoracic 4DCBCT reconstruction using projection data that was simulated using a clinical rapid 3DCBCT acquisition protocol. Methods We simulated conventional (1320 projections over 4 min) and rapid (491 projections over 16.6 s) CBCT acquisitions using 4D computed tomography (CT) volumes of 14 lung cancer patients. Conventional acquisition data were reconstructed using the 4D Feldkamp‐Davis‐Kress (FDK) algorithm. Rapid acquisition data were reconstructed using 3DFDK, 4DFDK, and Motion‐Compensated FDK (MCFDK). Image quality was evaluated using Contrast‐to‐Noise Ratio (CNR), Tissue Interface Width (TIW), Root‐Mean‐Square Error (RMSE), and Structural SIMilarity (SSIM). Results The conventional acquisition 4DFDK reconstructions had median phase averaged CNR, TIW, RMSE, and SSIM of 2.96, 8.02 mm, 83.5, and 0.54, respectively. The rapid acquisition 3DFDK reconstructions had median CNR, TIW, RMSE, and SSIM of 2.99, 13.6 mm, 112, and 0.44 respectively. The rapid acquisition MCFDK reconstructions had median phase averaged CNR, TIW, RMSE, and SSIM of 2.98, 10.2 mm, 103, and 0.46, respectively. Rapid acquisition 4DFDK reconstruction quality was insufficient for any practical use due to sparse angular projection sampling. Conclusions Results suggest that 4D motion‐compensated reconstruction of rapid acquisition thoracic CBCT data are feasible with image quality approaching conventional acquisition CBCT data reconstructed using standard 4DFDK.

Keywords: computed tomography; motion; rapid acquisition; motion compensated; acquisition

Journal Title: Journal of Applied Clinical Medical Physics
Year Published: 2023

Link to full text (if available)


Share on Social Media:                               Sign Up to like & get
recommendations!

Related content

More Information              News              Social Media              Video              Recommended



                Click one of the above tabs to view related content.