LAUSR.org creates dashboard-style pages of related content for over 1.5 million academic articles. Sign Up to like articles & get recommendations!

Comparison of SimoaTM and EllaTM to assess serum neurofilament‐light chain in multiple sclerosis

Photo from wikipedia

We compared SimoaTM and EllaTM immunoassays to assess serum neurofilament‐light chain levels in 203 multiple sclerosis patients from the OFSEP HD study. There was a strong correlation (ρ = 0.86,… Click to show full abstract

We compared SimoaTM and EllaTM immunoassays to assess serum neurofilament‐light chain levels in 203 multiple sclerosis patients from the OFSEP HD study. There was a strong correlation (ρ = 0.86, p < 0.0001) between both platforms. The EllaTM instrument overestimated values by 17%, but as the data were linear (p = 0.57), it was possible to apply a correction factor to EllaTM results. As for SimoaTM, serum neurofilament‐light chain levels measured by EllaTM were correlated with age and EDSS and were significantly higher in active multiple sclerosis, suggesting that these assays are equivalent and can be used in routine clinical practice.

Keywords: serum neurofilament; neurofilament light; multiple sclerosis; light chain

Journal Title: Annals of Clinical and Translational Neurology
Year Published: 2021

Link to full text (if available)


Share on Social Media:                               Sign Up to like & get
recommendations!

Related content

More Information              News              Social Media              Video              Recommended



                Click one of the above tabs to view related content.