Cleaning cannulated medical devices can be challenging for perioperative and sterile processing department personnel. We performed a laboratory experimental study to evaluate differences in cleaning effectiveness using either a back-and-forth… Click to show full abstract
Cleaning cannulated medical devices can be challenging for perioperative and sterile processing department personnel. We performed a laboratory experimental study to evaluate differences in cleaning effectiveness using either a back-and-forth or helical spinning brushing motion and the effect of rinsing the bristles at each reintroduction of the brush in the lumen. We also tested the lumen cleanliness after high-pressure water cleansing without brushing. We inspected the devices to determine whether visible soil remained, and we measured the amount of residual organic matter using adenosine triphosphate testing to determine cleaning method effectiveness. The results showed that rinsing the brush during cleaning decreased the amount of organic material that remained in the lumen. A helical spinning motion with brush rinsing at each reintroduction of the brush may be more effective than back-and-forth brushing with rinsing, but additional testing with a larger sample size is required to determine whether this result is replicable.
               