The electrochemical aptamer-based (E-AB) biosensor usually has a long reaction time when detecting thrombin. This work reports the design of an E-AB biosensor with dual recognition sites to quickly detect… Click to show full abstract
The electrochemical aptamer-based (E-AB) biosensor usually has a long reaction time when detecting thrombin. This work reports the design of an E-AB biosensor with dual recognition sites to quickly detect thrombin. Specifically, two specific recognition sites of thrombin were used to design three aptamer sequences (TBA-15, TBA-29 and TBA-U), followed by fabrication of corresponding sensors. First, we tested these three types of biosensors in tris buffer solution, and found that the response time of the TBA-U sensor to the same concentration of thrombin was about 2 hours, which is shorter than TBA-15 and TBA-29 sensors. Then, we also did the same test in 50 % diluted serum with 500 nM thrombin. The response time of the TBA-U sensor was about 2 hours, which is still faster than the 3 hours of TBA-15 sensor and the 5.5 hours for TBA-29 sensor. In addition, in terms of dynamic range and specificity, TBA-U has good performance.
               