PURPOSE Ultra-high dose rate irradiation, also known as FLASH, has been shown to improve the therapeutic ratio of radiation therapy (RT). The mechanism behind this effect has been partially explained… Click to show full abstract
PURPOSE Ultra-high dose rate irradiation, also known as FLASH, has been shown to improve the therapeutic ratio of radiation therapy (RT). The mechanism behind this effect has been partially explained by the radiochemical oxygen depletion (ROD) hypothesis, which attributes the protection of the normal tissue to the induction of transient hypoxia by ROD. To better understand the contribution of oxygen to the FLASH effect, it is necessary to measure oxygen (O2 ) in vivo during FLASH irradiation. This study's goal is to determine the temporal resolution required to accurately measure the rapidly changing oxygen concentration immediately after FLASH irradiation. METHODS We conducted a computational simulation of oxygen dynamics using a real vascular model that was constructed from a public fluorescence microscopy dataset. The dynamic distribution of oxygen tension (po2 ) during and after FLASH RT was modeled by a partial differential equation (PDE) considering oxygen diffusion, metabolism, and ROD. The underestimation of ROD due to oxygen recovery was evaluated assuming either complete or partial depletion, and a range of possible values for parameters such as oxygen diffusion, consumption, vascular po2 and vessel density. RESULT The O2 concentration recovers rapidly after FLASH RT. Assuming a temporal resolution of 0.5 s, the estimated ROD is only 50.7% and 36.7% of its actual value in cases of partial and complete depletion, respectively. Additionally, the underestimation of ROD is highly dependent on the vascular density. To estimate ROD rate with 90% accuracy, temporal resolution on the order of milliseconds is required considering the uncertainty in parameters involved, especially, the diverse vascular density of the tissue. CONCLUSION The rapid recovery of O2 poses a great challenge for in vivo ROD measurements during FLASH RT. Temporal resolution on the order of milliseconds is recommended for ROD measurements in the normal tissue. Further work is warranted to investigate whether the same requirements apply to tumors, given their irregular vasculature. This article is protected by copyright. All rights reserved.
               
Click one of the above tabs to view related content.