To protect patients from RF heating in MRI scan, this work proposes an accurate and patient‐specific whole‐body specific absorption rate (SAR) characterization method based on an equivalent circuit model. Compared… Click to show full abstract
To protect patients from RF heating in MRI scan, this work proposes an accurate and patient‐specific whole‐body specific absorption rate (SAR) characterization method based on an equivalent circuit model. Compared to the standard pulse energy method defined in NEMA MS 8‐2016, this method avoids the complexity of integrating flux loops and has the potential to be easily implemented in MRI scanners.
               