Accuracy of volume and pressure delivery by mechanical ventilators in use in neonatal intensive care units: A quality control study

Photo from wikipedia

Despite technical specifications of neonatal mechanical ventilators (MVs) guarantee clinically irrelevant discrepancies between the set and the delivered values of ventilation parameters, previous studies reported large deviations. Most studies characterized… Click to show full abstract

Despite technical specifications of neonatal mechanical ventilators (MVs) guarantee clinically irrelevant discrepancies between the set and the delivered values of ventilation parameters, previous studies reported large deviations. Most studies characterized performances of a given model/brand by studying a single device, disregarding possible intramodel differences, and leaving the accuracy of the ventilation parameters effectively delivered in clinical settings unknown. The aim of this study was to evaluate the realā€life accuracy of pressure and volume parameters delivered by neonatal ventilators ready to be used on patients in neonatal intensive care units (NICUs).

Keywords: care units; mechanical ventilators; intensive care; accuracy; pressure; neonatal intensive

Journal Title: Pediatric Pulmonology
Year Published: 2020