Microfluidic immunoassay devices are a promising technology that can quickly detect biomarkers with high sensitivity. Recently, many studies implementing this technology on paper substrates have been proposed for improving cost… Click to show full abstract
Microfluidic immunoassay devices are a promising technology that can quickly detect biomarkers with high sensitivity. Recently, many studies implementing this technology on paper substrates have been proposed for improving cost and user-friendliness. However, these studies have identified problems with the large volume of sample required, low sensitivity, and a lack of quantitative accuracy and precision. In this paper, we report a novel structure implemented as a cellulosic material-based microchannel device capable of quantitative immunoassay using small sample volumes. We fabricated microfluidic channels between a transparent cellophane film and water-resistant paper to facilitate loading of small-volume samples and reagents, with a 40-μm-wide immunoreaction matrix constructed in the center of the microchannel using highly precise photolithography. A fluorescence sandwich immunoassay for C-reactive protein (CRP) was successfully implemented that required only a 1-μL sample volume and a 20-min reaction time. We confirmed that the limit of detection of the device was 10–20 ng/mL with a coefficient of variation under 5.6%, which is a performance level comparable to conventional plastic-based human CRP enzyme-linked immunosorbent assay (ELISA) kits. We expect that such devices will lead to the elimination of large amounts of medical waste from the use of ubiquitous diagnostics, a result that is consistent with environmental sustainability goals.
               