The analysis of tobramycin was demonstrated successfully as an example for electrospray ionization on an open-source hardware ion mobility spectrometer. This instrument was assembled inexpensively in-house, and required only very… Click to show full abstract
The analysis of tobramycin was demonstrated successfully as an example for electrospray ionization on an open-source hardware ion mobility spectrometer. This instrument was assembled inexpensively in-house, and required only very few purpose-made components. The quantitative determination of tobramycin required 20 s for a reading. The calibration curve for the range from 50 to 200 μM was found to be linear with a correlation coefficient of r = 0.9994. A good reproducibility was obtained (3% relative standard deviation) and the limit of detection was determined as 8 μM. As the concentration of the active ingredient in the eye drops (ophthalmic solutions) is too high for the sensitivity of the instrument, the samples had to be diluted appropriately.
               
Click one of the above tabs to view related content.