The electroporation of cells is nowadays used for a large variety of purposes, from basic research to cancer therapy and food processing. Understanding molecular mechanisms of the main processes involved… Click to show full abstract
The electroporation of cells is nowadays used for a large variety of purposes, from basic research to cancer therapy and food processing. Understanding molecular mechanisms of the main processes involved in electroporation is thus of significant interest. In the present work, we propose an experimental system to record in real time the evolution of any cell parameter which can be evaluated by fluorescence (before, during and after application of the electroporation pulses to cells in suspension). The system is based on the design of adequate electroporation electrodes, compatible with a standard spectrofluorometer cuvette housing. The electric field intensity generated when pulses are applied was carefully characterized for different geometries of the electrodes, to choose a construction ensuring the greatest homogeneity of the field in combination with the best possible illumination of the sample. As an example of the method’s application, we present here generalized polarization kinetics for a varying number of electroporation pulses applied to a cell suspension; the general polarization parameter is strongly correlated to water presence in the hydrophobic membrane core. The system may be used for many other fluorescence measurements useful for the characterization of the electroporation process.
               