LAUSR.org creates dashboard-style pages of related content for over 1.5 million academic articles. Sign Up to like articles & get recommendations!

Prosthetic liner wear in total hip replacement: a longitudinal 13-year study with computed tomography

Photo from wikipedia

This case report follows a woman who had a total hip replacement in 1992 when she was 45 years old. Six serial computed tomography (CT) examinations over a period of 13 years… Click to show full abstract

This case report follows a woman who had a total hip replacement in 1992 when she was 45 years old. Six serial computed tomography (CT) examinations over a period of 13 years provided information that allowed her revision surgery to be limited to liner replacement as opposed to replacement of the entire prosthesis. Additionally, they provided data that ruled out the presence of osteolysis and indeed none was found at surgery. In 2004, when the first CT was performed, the 3D distance the femoral head had penetrated into the cup was determined to be 2.6 mm. By 2017, femoral head penetration had progressed to 5.0 mm. The extracted liner showed wear at the thinnest part to be 5.5 mm, as measured with a micrometer. The use of modern CT techniques can identify problems, while still correctable without major surgery. Furthermore, the ability of CT to assess the direction of wear revealed that the liner wear changed from the cranial to dorsal direction.

Keywords: replacement; computed tomography; liner wear; hip replacement; total hip

Journal Title: Skeletal Radiology
Year Published: 2018

Link to full text (if available)


Share on Social Media:                               Sign Up to like & get
recommendations!

Related content

More Information              News              Social Media              Video              Recommended



                Click one of the above tabs to view related content.