Stent implantation is a common procedure followed in arteries affected by atherosclerosis. This procedure can lead to other stenting-related problems. One of these is the deposition and accumulation of blood… Click to show full abstract
Stent implantation is a common procedure followed in arteries affected by atherosclerosis. This procedure can lead to other stenting-related problems. One of these is the deposition and accumulation of blood clots over stent struts. This process can have further consequences, in so far as it can introduce modifications to the flow pattern. This problem is especially critical in stented bifurcations, where resulting stent geometry is more complex. In this regard, a numerical study is presented of the effect on the flow pattern and platelet activation of blood clot depositions on the stent struts of a stented coronary bifurcation. The numerical model is first validated with experimental measurements performed for this purpose. Experiments considered a flow with suspended artificial thrombi, which naturally deposited on stent struts. The location and shape observed were used to create numerical thrombi. Following this, numerical simulations were performed to analyze the influence of the presence of thrombi depositions on parameters such as Time Averaged Wall Shear Stress, Oscillatory Shear Index or Relative Residence Time. Finally, a study was also carried out of the effect of different geometrical configurations, from a straight tube to a stented bifurcation model with thrombus depositions, on platelet activation.
               