The development of technologies for using the Novac-11 pulsed electron accelerator in radiation therapy of animals with spontaneous neoplasms requires dosimetric and radiobiological studies. The studies were performed on cultured… Click to show full abstract
The development of technologies for using the Novac-11 pulsed electron accelerator in radiation therapy of animals with spontaneous neoplasms requires dosimetric and radiobiological studies. The studies were performed on cultured Chinese hamster V-79 fibroblasts after irradiation with 10 MeV electrons in a dose range up to 12 Gy and 60Co γ-radiation. Chemical dosimeters FBX and Fricke were used as additional test-systems. The depth dose curves were measured and the maximum dose depth of the electron beam was determined in tissue-equivalent phantoms. Cell survival and the data of chemical dosimetric systems showed that the effects of electron irradiation did not differ from that of 60Co γ-radiation. It was concluded that the use of Novac-11 in the therapy of animals with spontaneous neoplasms is advisable.
               