Reproducibility of WC–C High-Temperature Fixed Point

Photo by fabiooulucas from unsplash

Reproducibility of the tungsten carbide–carbon peritectic (WC–C) fixed point (3021 K) was investigated by comparing six WC–C blackbody-type cells with each other. All the cells were built at VNIIOFI and had… Click to show full abstract

Reproducibility of the tungsten carbide–carbon peritectic (WC–C) fixed point (3021 K) was investigated by comparing six WC–C blackbody-type cells with each other. All the cells were built at VNIIOFI and had the same design with a cell outer diameter of 24 mm and a cavity opening diameter of 3  mm. Four cells were built using tungsten powder supplied by Alfa Aesar, while the other two cells used powder from Zhuzhou KETE (China). The nominal purity of both suppliers was 99.999 %. All the cells were compared in the same furnace. A reference cell in the second furnace was used for monitoring the stability of a radiation thermometer. Melting temperatures (given by the point of inflection) of all six cells agreed within ±45 mK. The reproducibility, as a standard deviation of the measured temperatures, can be estimated as 35 mK. The Zhuzhou KETE cells showed a slightly lower temperature than the Alfa Aesar cells: by 32 mK on average, which is comparable with repeatability of the measurements.

Keywords: reproducibility high; high temperature; reproducibility; point; fixed point

Journal Title: International Journal of Thermophysics
Year Published: 2017

Link to full text (if available)


Share on Social Media:                               Sign Up to like & get
recommendations!

Related content

More Information              News              Social Media              Video              Recommended



                Click one of the above tabs to view related content.