X-ray scatter is a major cause of image quality degradation in dimensional CT. Especially, in case of highly attenuating components scatter-to-primary ratios may easily be higher than 1. The corresponding artifacts which appear as cupping or dark streaks in the CT reconstruction may impair a metrological assessment. Therefore, an appropriate scatter correction is crucial. Thereby, the gold standard is to predict the scatter distribution using a Monte Carlo (MC) code and subtract the corresponding scatter estimate from the measured raw data. MC, however, is too slow to be used routinely. To correct for scatter in real-time, we developed the deep scatter estimation (DSE). It uses a deep convolutional neural network which is trained to reproduce the output of MC simulations using only the acquired projection data as input. Once trained, DSE can be applied in real-time. The present study demonstrates the potential of the proposed approach using simulations and measurements. In both cases the DSE yields highly accurate scatter estimates that differ by< 3% from our MC scatter predictions. Further, DSE clearly outperforms kernel-based scatter estimation techniques and hybrid approaches, as they are in use today.
               