A simple correction method for isobaric interferences induced by lead during uranium isotope analysis using secondary ion mass spectrometry

Photo from wikipedia

Isobaric interference is a major limitation of secondary ion mass spectrometry. We developed a simple correction method for polyatomic mass interferences from lead in isotope ratio measurements of uranium. Lead-generated… Click to show full abstract

Isobaric interference is a major limitation of secondary ion mass spectrometry. We developed a simple correction method for polyatomic mass interferences from lead in isotope ratio measurements of uranium. Lead-generated isobars were measured to determine their formation rates relative to lead isotopes. The rates were used to mathematically subtract the isobar contributions from the signal intensities in the uranium mass range. The correction method was successfully verified using mixed uranium-lead samples (oxide powder and solution-dried residue).

Keywords: secondary ion; ion mass; correction method; mass

Journal Title: Journal of Radioanalytical and Nuclear Chemistry
Year Published: 2018