Organophosphorus pesticides (OPs) are widely used in agriculture and the monitoring of their residues is very important to protect human health. Immunoassays are important tools for the analysis of small… Click to show full abstract
Organophosphorus pesticides (OPs) are widely used in agriculture and the monitoring of their residues is very important to protect human health. Immunoassays are important tools for the analysis of small molecules. Generally, noncompetitive mode of immunoassay is considered to be more sensitive than competitive mode. In this study, peptides that can identify immunocomplex of OPs were screened from a phage display library. Subsequently, a second-generation peptide library was constructed and peptides with better performance were isolated. Then, a rapid and sensitive noncompetitive magnetic-phage anti-immunocomplex assay (MPHAIA) for OPs was developed based on the best phage-peptide and single chain antibody immunomagnetic beads. The MPHAIA showed broad specificity for OPs with a thiophosphate group. The half-saturated concentration (SC50) values and limits of detection (LODs) of MPHAIA to 12 OPs were ranged from 15.04 to 105.48 ng/mL and 4.07-14.19 ng/mL, respectively. The accuracy and reliability of MPHAIA were verified by gas chromatography-tandem mass spectrometry (GC-MS/MS) parallel analysis of six kinds of OPs in spiked cucumber samples. The recovery rates were in range of 81.2-116.3% with coefficient of variation from 4.1% to 14.1%, which were consistent with the results of GC-MS/MS.
               