LAUSR.org creates dashboard-style pages of related content for over 1.5 million academic articles. Sign Up to like articles & get recommendations!

Characterization of a new candidate isotopic reference material for natural Pb using primary measurement method.

Photo by bondomovies from unsplash

A lead isotopic standard solution with natural abundance has been developed by applying a mixture of a solution of enriched 208Pb and a solution of enriched 204Pb (208Pb-204Pb double spike… Click to show full abstract

A lead isotopic standard solution with natural abundance has been developed by applying a mixture of a solution of enriched 208Pb and a solution of enriched 204Pb (208Pb-204Pb double spike solution) as bracketing method. The amount-of-substance ratio of 208Pb:204Pb in this solution is accurately measured by applying EDTA titrimetry, which is one of the primary measurement methods, to each enriched Pb isotope solution. Also metal impurities affecting EDTA titration and minor lead isotopes contained in each enriched Pb isotope solution are quantified by ICP-SF-MS. The amount-of-substance ratio of 208Pb:204Pb in the 208Pb-204Pb double spike solution is 0.961959 ± 0.000056 (combined standard uncertainty; k = 1). Both the measurement of lead isotope ratios in a candidate isotopic standard solution and the correction of mass discrimination in MC-ICP-MS are carried out by coupling of a bracketing method with the 208Pb-204Pb double spike solution and a thallium internal addition method, where thallium solution is added to the standard and the sample. The measured lead isotope ratios and their expanded uncertainties (k = 2) in the candidate isotopic standard solution are 18.0900 ± 0.0046 for 206Pb:204Pb, 15.6278 ± 0.0036 for 207Pb:204Pb, 38.0626 ± 0.0089 for 208Pb:204Pb, 2.104406 ± 0.00013 for 208Pb:206Pb, and 0.863888 ± 0.000036 for 207Pb:206Pb. The expanded uncertainties are about one half of the stated uncertainty for NIST SRM 981, for 208Pb:204Pb, 207Pb:204Pb and 206Pb:204Pb, or one eighth, for 208Pb:206Pb and 207Pb:206Pb, The combined uncertainty consists of the uncertainties due to lead isotope ratio measurements and the remaining time-drift effect of mass discrimination in MC-ICP-MS, which is not removed by the coupled correction method. In the measurement of 208Pb:204Pb, 207Pb:204Pb and 206Pb:204Pb, the latter contribution is two or three times larger than the former. When the coupling of a bracketing method with the 208Pb-204Pb double spike solution and a thallium internal addition method is applied to the analysis of NIST SRM 981, the measured lead isotope ratios are in good agreement with its certified values. This proves that the developed method is not only consistent with the conventional one by NIST SRM 981 but also enables measurement of the lead isotope ratios with higher precision.

Keywords: method; 208pb 204pb; lead isotope; solution; candidate isotopic

Journal Title: Analytica chimica acta
Year Published: 2017

Link to full text (if available)


Share on Social Media:                               Sign Up to like & get
recommendations!

Related content

More Information              News              Social Media              Video              Recommended



                Click one of the above tabs to view related content.