Aspiration Thrombectomy-Assisted Endovascular Retrieval of an Embolized Angio-Seal Device Causing Claudication.

Photo by nilsynils from unsplash

The Angio-Seal vascular closure device is used to reduce time to hemostasis after femoral artery puncture. Although rare, complications associated with Angio-Seal are significant, including infection, pseudoaneurysm formation, symptomatic femoral… Click to show full abstract

The Angio-Seal vascular closure device is used to reduce time to hemostasis after femoral artery puncture. Although rare, complications associated with Angio-Seal are significant, including infection, pseudoaneurysm formation, symptomatic femoral artery stenosis, and separation and embolization, leading to limb-threatening ischemia. This report describes Angio-Seal polymer anchor embolization to the tibioperoneal trunk successfully retrieved using the 8-French Indigo Aspiration System.

Keywords: aspiration thrombectomy; thrombectomy assisted; angio seal; assisted endovascular; device