Relationship between the location of the ventricular catheter tip and the ventriculoperitoneal shunt malfunction

Photo from wikipedia

OBJECTIVE Ventriculoperitoneal (VP) shunt is the most common neurosurgical procedure to treat obstructive and communicating hydrocephalus, but failures are frequent. The most common causes of shunt failure are malpositioning and… Click to show full abstract

OBJECTIVE Ventriculoperitoneal (VP) shunt is the most common neurosurgical procedure to treat obstructive and communicating hydrocephalus, but failures are frequent. The most common causes of shunt failure are malpositioning and obstruction of the ventricular catheter by debris or blood clothes. The knowledge of the relationship between ventricular catheter tip position, etiology of hydrocephalus and patient's age with shunt malfunction may be useful to avoid shunt failure. PATIENTS AND METHODS We retrospectively analyze 89 adult patients affected by obstructive and communicating hydrocephalus operated with Ventriculoperitoneal shunt at our Institute. Patients with evident abdominal shunt malfunction were excluded from this study. Statistical analysis was performed in the group of patients with intracranial catheter malfunction in order to correlate shunt malfunction with the position of the catheter tip in brain ventricles, etiology of hydrocephalus and patient's age. RESULTS Shunt revision was performed in 26 patients out of 89 and cranial catheter malfunction was documented in 11 patients out of 26. Tip position in Monro foramen, lateral ventricles and third ventricle ("good position") had a lower rate of surgical shunt revision compared to septum pellucidum and wall of lateral ventricle ("bad position") p value = 0,049. No statistical significant association was observed between shunt malfunction and etiology of the hydrocephalus or patient's age. CONCLUSIONS Ventricular catheter tip position is one the most important factor for shunt function.

Keywords: etiology; malfunction; shunt malfunction; ventricular catheter; shunt

Journal Title: Clinical Neurology and Neurosurgery
Year Published: 2018

Link to full text (if available)


Share on Social Media:                               Sign Up to like & get
recommendations!

Related content

More Information              News              Social Media              Video              Recommended



                Click one of the above tabs to view related content.