OBJECTIVE To assay EEG signal quality recorded with tripolar concentric ring electrodes (TCREs) compared to regular EEG electrodes. METHODS EEG segments were recorded simultaneously by TCREs and regular electrodes, low-pass… Click to show full abstract
OBJECTIVE To assay EEG signal quality recorded with tripolar concentric ring electrodes (TCREs) compared to regular EEG electrodes. METHODS EEG segments were recorded simultaneously by TCREs and regular electrodes, low-pass filtered at 35 Hz (REG35) and 70 Hz (REG70). Clips were rated blindly by nine electroencephalographers for presence or absence of key EEG features, relative to the "gold-standard" of the clinical report. RESULTS TCRE showed less EMG artifact (F = 15.4, p < 0.0001). Overall quality rankings were not significantly different. Focal slowing was better detected by TCRE and spikes were better detected by regular electrodes. Seizures (n = 85) were detected by TCRE in 64 cases (75.3%), by REG70 in 75 (88.2%) and REG35 in 69 (81.2%) electrodes. TCRE detected 9 (10.6%) seizures not detected by one of the other 2 methods. In contrast, 14 seizures (16.5%) were not detected by TCRE, but were by REG35 electrodes. Each electrode detected interictal spikes when the other did not. CONCLUSIONS TCRE produced similar overall quality and confidence ratings versus regular electrodes, but less muscle artifact. TCRE recordings detected seizures in 7% of instances where regular electrodes did not. SIGNIFICANCE The combination of the two types increased detection of epileptiform events compared to either alone.
               