LAUSR.org creates dashboard-style pages of related content for over 1.5 million academic articles. Sign Up to like articles & get recommendations!

Design of a tritium gas cell for beta-ray induced X-ray spectrometry using Monte Carlo simulation

Photo from wikipedia

Abstract One of the methods used for tritium gas analysis is beta-ray induced X-ray spectrometry (BIXS). Gas cell design is important in this method. The structure of the gas cell… Click to show full abstract

Abstract One of the methods used for tritium gas analysis is beta-ray induced X-ray spectrometry (BIXS). Gas cell design is important in this method. The structure of the gas cell for BIXS was optimized by Monte Carlo simulation of beta-ray induced X-ray spectra in various window geometries using the Geant4 tool kit (version 10.01.p02). The simulated spectrum from tritium decay fitted the observed one, and the simulation model was used to obtain the cell parameters for BIXS. The optimum thickness of the gold layer on a beryllium window was around 150 nm. This simulation model also considered the relationship between self-absorption by hydrogen gas and the cell length. Self-absorption increased with increasing cell length and the relationship between the sample pressure and cell length was formulated.

Keywords: beta ray; ray induced; gas; simulation; cell; gas cell

Journal Title: Fusion Engineering and Design
Year Published: 2017

Link to full text (if available)