LAUSR.org creates dashboard-style pages of related content for over 1.5 million academic articles. Sign Up to like articles & get recommendations!

Modeling of the response function of CdTe detectors for a hard X-ray synthetic diagnostic

Photo from wikipedia

Abstract The response function of CdTe detectors used in a hard X-ray spectrometer system (HXRS) on the TCV tokamak fusion experiment is modeled based on the measured energy spectra from… Click to show full abstract

Abstract The response function of CdTe detectors used in a hard X-ray spectrometer system (HXRS) on the TCV tokamak fusion experiment is modeled based on the measured energy spectra from multiple gamma-ray sources. The response function consists of an asymmetric Gaussian photopeak, an exponentially inclined shelf structure below the photopeak, and CdTe escape peaks. It is demonstrated that the measured energy spectra of radioactive sources can be correctly reconstructed using this response function. This model is applied to a hard X-ray synthetic diagnostic used to convert a modeled electron distribution function to a measured hard X-ray signal, for the study of energetic electrons in the TCV tokamak.

Keywords: hard ray; function; ray; cdte; response function

Journal Title: Fusion Engineering and Design
Year Published: 2020

Link to full text (if available)


Share on Social Media:                               Sign Up to like & get
recommendations!

Related content

More Information              News              Social Media              Video              Recommended