Background The performance of the Roche Elecsys® Anti-SARS-CoV-2, Abbott Architect SARS-CoV-2 IgM and IgG, Euroimmun SARS-CoV-2 IgA, Euroimmun SARS-CoV-2 IgG ELISA, and Trillium IgG/IgM rapid assays was evaluated in Jamaica. Methods Diagnostic sensitivities of the assays were assessed by testing serum samples from SARS-CoV-2 PCR-confirmed persons and diagnostic specificity was assessed by testing serum samples collected during 2018-2019 from healthy persons and from persons with antibodies to a wide range of viral infections. Results Serum samples collected ≥14 days after onset of symptoms, or an initial SARS-CoV-2 RT-PCR positive test for asymptomatics, showed diagnostic sensitivities ranging from 67.9-75.0% when including all possible disease severities and increased to 90.0-95.0% when examining those with moderate to critical disease. Grouping moderate to critical disease showed a significant association with a SARS-CoV-2 antibody positive result for all assays. Diagnostic specificity ranged from 96.7-100.0%. For all assays examined, SARS-CoV-2 real-time PCR cycle threshold (Ct) values of the initial nasopharyngeal swab sample testing positive were significantly different for samples testing antibody positive versus negative. Conclusions These data from a predominantly African descent Caribbean population shows comparable diagnostic sensitivities and specificities for all testing platforms assessed and limited utility of these tests for persons with asymptomatic and mild infections.
               
Click one of the above tabs to view related content.