Adverse reaction to metal debris (ARMD) is an issue in metal-on-metal (MoM) total hip replacements (THR). It mainly affects large-head MoM THR, whereas 28-32 mm MoM pairings are associated with low… Click to show full abstract
Adverse reaction to metal debris (ARMD) is an issue in metal-on-metal (MoM) total hip replacements (THR). It mainly affects large-head MoM THR, whereas 28-32 mm MoM pairings are associated with low long-term revision rates. However, the bearing surface is not necessarily the only cause of metal debris. This report documents with advanced analysis of the retrievals a particular cause of trunnionosis in late failure of a small diameter MoM THR and illustrates the importance of cleaning of the taper when seating the head in THR. A 65-year-old patient was revised due to ARMD 16 years after small diameter MoM THR. Debridement and exchange of the inlay and the head had been performed through an anterior approach. While the cup and the outer surface of the head were accessible to direct analysis by an optical coordinate measuring machine, the female taper had to be analysed indirectly by measuring an imprint. Wear from the cup and the head was within expected low ranges. The analysis of the female taper identified bone fragments, which contributed to trunnionosis. Failure due to ARMD after MoM THR is not necessarily caused by the bearing, but can be due to trunnionosis. Bone fragments within the taper contact in this case highlight the importance of meticulous cleaning of the taper before seating the head, to avoid trunnionosis.
               