Abstract A new evaporative light scattering detector (ELSD) for the analysis of polyolefins by high temperature gel permeation chromatography (GPC) was recently introduced by Agilent Technologies. For the first time,… Click to show full abstract
Abstract A new evaporative light scattering detector (ELSD) for the analysis of polyolefins by high temperature gel permeation chromatography (GPC) was recently introduced by Agilent Technologies. For the first time, we investigated the possibility to use this detector to measure the molecular weight distributions (MWD) of different types of polyolefins (polypropylene, linear and low-density polyethylene) in dibutoxymethane (DBM, butylal). These samples were previously characterized by GPC in trichlorobenzene (TCB) with a differential refractive index (DRI) detector in an interlaboratory study conducted by the International Union of Pure and Applied Chemistry (IUPAC) [1], and in a recent publication by GPC with the new ELSD in TCB [2]. The signal to noise of ELSD using DBM is about 10 times lower than that for TCB. However, the ELSD signal power exponent for DBM was measured as 1.35, which is much closer to unity than the value of 1.61 for TCB. After applying the required corrections to linearize the response of the ELSD signal as a function of concentration, similar average molecular weights to those measured in the interlaboratory study using DRI, were obtained for the analyzed resins.
               
Click one of the above tabs to view related content.