Abstract The energy distribution of an electron gun is one of the most important characteristics determining the performance of electron beam-based instruments, such as electron microscopes and electron energy loss… Click to show full abstract
Abstract The energy distribution of an electron gun is one of the most important characteristics determining the performance of electron beam-based instruments, such as electron microscopes and electron energy loss spectroscopes. For accurate measurements of the energy distribution, this study presents a novel retarding field energy analyzer (RFEA) with the feature of an additional integrated pre-lens, which enables an adjustment of beam trajectory into the analyzer. The advantages of this analyzer are its compact size and simple electrode configuration. According to trajectory simulation theories, the optimum condition arises when the incident electron beam inside the RFEA is focused on the center of a retarding electrode. Comparing I–V curves depending on whether the pre-lens working or not, it is confirmed that the use of the pre-lens dramatically improves the energy resolution and efficiency of the signal acquisition process. The pre-lens RFEA was applied to characterize a Schottky electron gun under various temperatures and extraction voltages as operational conditions. When the tip temperature was increased by 50 K, we were able to measure an energy distribution broadening of 13.8 meV with the proposed pre-lens RFEA. The relative standard deviation of energy distribution was 0.7% for each working condition.
               