In this study, for the first time, the automated digestion and sequencing of an RNA molecule via the use of immobilized RNase cartridges attached to a multidimensional liquid chromatography (LC)-mass… Click to show full abstract
In this study, for the first time, the automated digestion and sequencing of an RNA molecule via the use of immobilized RNase cartridges attached to a multidimensional liquid chromatography (LC)-mass spectrometry (MS) system are presented. We first developed an on-line digestion-HILIC two-dimensional (2D)-LC-MS method in order to sequence CRISPR guide RNAs for gene editing. Three RNases (T1, A, and U2) were immobilized on polyetheretherketone cartridges, and their performance was evaluated. Ultrafast digestions were performed within 2.3 min with the on-line approach versus 30 min via the conventional off-line approach. The higher sequence coverage was achieved by the RNase T1 (71%), which is the same as the off-line mode. A 20-fold reduction in the gRNA sample amount was achieved with the on-line digestion approach (6.5 μg) in comparison to that with the off-line approach (130 μg). In the second step, a three-dimensional (3D)-LC-MS method was developed for the sequencing of fractions collected on-line across the main peak and the partially separated tail by the reference ion-pairing RPLC method. Additional insights were gained in order to better understand the cause of the main peak tailing.
               