LAUSR.org creates dashboard-style pages of related content for over 1.5 million academic articles. Sign Up to like articles & get recommendations!

How to Improve the Resolving Power of Compact Electrospray Ionization Ion Mobility Spectrometers

Photo from wikipedia

Every drift tube ion mobility spectrometer (IMS) has an optimum drift voltage to reach maximum resolving power. This optimum depends, among other things, on the temporal and spatial width of… Click to show full abstract

Every drift tube ion mobility spectrometer (IMS) has an optimum drift voltage to reach maximum resolving power. This optimum depends, among other things, on the temporal and spatial width of the injected ion packet and the pressure within the IMS. A reduction of the spatial width of the injected ion packet leads to improved resolving power, higher peak amplitudes when operating the IMS at optimum resolving power, and thus a better signal-to-noise ratio despite the reduced number of injected ions. Hereby, the performance of electrospray ionization (ESI)-IMS can be considerably improved. By setting the ion shutter opening time to just 5 μs and slightly increasing the pressure, a high resolving power RP > 150 can be achieved with a given drift length of just 75 mm. At such high resolving power, even a mixture of the herbicides isoproturon and chlortoluron having similar ion mobility can be well separated despite short drift length.

Keywords: electrospray ionization; resolving power; ion; ion mobility

Journal Title: Analytical Chemistry
Year Published: 2023

Link to full text (if available)


Share on Social Media:                               Sign Up to like & get
recommendations!

Related content

More Information              News              Social Media              Video              Recommended



                Click one of the above tabs to view related content.