A simple and fast pre-column derivatization high performance liquid chromatographic method for the technical analysis of mancozeb was developed and validated in the present study. The mobile phase consists of… Click to show full abstract
A simple and fast pre-column derivatization high performance liquid chromatographic method for the technical analysis of mancozeb was developed and validated in the present study. The mobile phase consists of a mixture of acetonitrile and 0.1% (v/v) formic acid in the proportion 60ā:ā40. This was found to give a sharp peak of the S-ethyl derivative of mancozeb (S-ethyl ethylene bis-dithiocarbamate) at a retention time of 9.12 minutes. HPLC analysis of mancozeb was carried out at a wavelength of 272 nm with a flow rate of 1.0 mL per minute. The linear regression analysis data for the calibration curve showed a good linear relationship with a regression coefficient of 0.999 in the concentration range of 8.2 mg Lā1 to 32.3 mg Lā1 for the derivative of mancozeb technical. The method was validated for specificity, linearity, precision and accuracy.
               