Precise positioning of magnetic particles and magnetized cells in lab-on-a-chip systems has attracted broad attention. Recently, drawing inspiration from electrical circuits, we have demonstrated a magnetic particle transport platform composed… Click to show full abstract
Precise positioning of magnetic particles and magnetized cells in lab-on-a-chip systems has attracted broad attention. Recently, drawing inspiration from electrical circuits, we have demonstrated a magnetic particle transport platform composed of patterned magnetic thin films in a microfluidic environment, which accurately moves the particles and single cells to specific spots, called capacitors. However, we have made no prior attempts to optimize the capacitor geometry. Here, we carefully analyze various design parameters and their effect on capacitor operation. We run simulations based on finite element methods and stochastic numerical analysis using our semi-analytical model. We then perform the required experiments to study the loading efficiency of capacitors with different geometries for magnetic particles of multiple sizes. Our experimental results agree well with the design criteria we developed based on our simulation results. We also show the capability of designed capacitors in storing the magnetically labeled cells and illustrate using them in a pilot drug screening application. These results are directly applicable to the design of robust platforms capable of transporting and assembling a large number of single particles and single cells in arrays, which are useful in the emerging field of single-cell analysis.
               
Click one of the above tabs to view related content.