LAUSR.org creates dashboard-style pages of related content for over 1.5 million academic articles. Sign Up to like articles & get recommendations!

Determination of Copper(II) Through Anodic Stripping Voltammetry in Tartrate Buffer Using an Antimony Film Screen-printed Carbon Electrode

Photo by seemurray from unsplash

ABSTRACT The voltammetric performance of an in situ plated antimony film screen-printed carbon electrode in hydrochloric acid, acetate buffer, and tartrate buffer was evaluated for the detection of copper(II) with… Click to show full abstract

ABSTRACT The voltammetric performance of an in situ plated antimony film screen-printed carbon electrode in hydrochloric acid, acetate buffer, and tartrate buffer was evaluated for the detection of copper(II) with differential pulse anodic stripping voltammetry. The tartrate buffer was superior, providing high sensitivity and good separation of copper and antimony stripping peaks. The analytical conditions for the determination of copper(II) were optimized. The detection limit was estimated to be 0.14 µg L−1 copper(II) and the relative standard deviation for 2.5 µg L−1 copper(II) was 3%. The applicability of the method was illustrated by the analysis of soil conditioner samples.

Keywords: tartrate buffer; antimony film; copper; film screen; buffer

Journal Title: Analytical Letters
Year Published: 2017

Link to full text (if available)


Share on Social Media:                               Sign Up to like & get
recommendations!

Related content

More Information              News              Social Media              Video              Recommended



                Click one of the above tabs to view related content.