ABSTRACT This work proposes a flow injection analysis system for sulfaguanidine determination in pharmaceutical and food samples. The method was based on the reaction of sulfaguanidine with nitronium ion to… Click to show full abstract
ABSTRACT This work proposes a flow injection analysis system for sulfaguanidine determination in pharmaceutical and food samples. The method was based on the reaction of sulfaguanidine with nitronium ion to produce a colored complex whose absorbance was measured at 545 nm. The flow injection analysis system’s significant parameters were checked by a fractional factorial design 27–2 and optimization by a Doehlert matrix. The flow injection analysis system shows optimum values at 0.28, 2.00, and 0.11% (w/v) for N-naphtil ethylenediamine, ammonium sulfamate, and sodium nitrite concentrations, respectively. The possible interferents present in pharmaceutical and food samples were assessed by a multivariate technique and depicted on probability charts, indicating no significant interferences at the 95% level of confidence interval. The method showed detection and quantification limits of 0.012 and 0.039 mg L−1, an analytical frequency of 30 readings h−1 and precision always lower than 5.0% expressed as the relative standard deviation. The obtained results were in accordance with those obtained by high-performance liquid chromatography.
               