A comparison of the analysis of respirable crystalline silica direct-on-filter methods using X-ray diffraction (XRD) and Fourier transform infrared (FT-IR) spectroscopy was undertaken using 253 real workplace air samples from road construction and tunnelling, coal mining, and kitchen benchtop manufacturing in Australia. Using pure α-quartz standards, XRD and FT-IR direct-on-filter analyses produced identical test results, however, the real workplace samples showed a clear discrepancy between FT-IR and XRD results with on average a 9% positive bias of the FT-IR results. The cause of the positive bias was due to matrix interferences which was confirmed by using synthetic mixture air samples. Approximately a third of the data by direct-on-filter method using FT-IR was assessed to be invalid based on the peak height ratio criterion due to excessive interferences and weight overload limitations. The XRD method showed better results due to less interference from the common matrices. XRD could handle up to twice the sample loading and at higher loadings up to 7 mg when a correction was applied. It was also able to achieve a lower limit of detection of 2 µg filter-1 when a slower scan condition was utilized.
               