Phantom Study of the Entrance Surface Dose of a New CT Scout Acquisition that also Serves as a Tube Warm-Up.

Photo by chernus_tr from unsplash

We determined the effect of a new scout image acquisition technique ('smart scout'), which also serves as a tube warm-up, on radiation dose and automatic exposure control (AEC) mA settings.… Click to show full abstract

We determined the effect of a new scout image acquisition technique ('smart scout'), which also serves as a tube warm-up, on radiation dose and automatic exposure control (AEC) mA settings. The entrance surface dose (ESD) of a chest phantom with and without the smart scout was measured. A conical AEC phantom was scanned in the setting for abdominal CT, and AEC curves were generated. ESD when the smart scout was not used was 0.75 mGy at 120 kV, 50 mA. ESD when using the smart scout was 0.24 mGy for a body habitus setting of 'Less', 0.54 mGy for 'Moderate' and 0.95 mGy for 'More'. When the diameter of the subject was ≥32 cm, the mA setting became lower in 'Less'. The smart scout reduced exposure at the 'Less' and Moderate' settings compared to the conventional scout scan.

Keywords: phantom; scout; serves tube; smart scout; new scout; also serves

Journal Title: Radiation protection dosimetry
Year Published: 2022