A proof-of-concept study of a new detector based on a thin plastic scintillator monitored by a Charge-Coupled Device (CCD) camera designed for monitoring and characterisation of Linac photon beams is… Click to show full abstract
A proof-of-concept study of a new detector based on a thin plastic scintillator monitored by a Charge-Coupled Device (CCD) camera designed for monitoring and characterisation of Linac photon beams is presented. The response of the detector is compared with radiochromic film using 6 and 18 MV radiotherapeutic beams. We have observed: (i) all instruments survived the secondary radiation fields during Linac operation, (ii) it was possible to process the measured data using statistical techniques and (iii) the processed data from the CCD camera qualitatively correspond to film dosimetry results. A statistical technique based on the selection of minimal values provides the clearest results. Quantitatively, CCD and film results can only be compared as 6 to 18 MV response rates. We have observed that the rates from the CCD data are systematically higher than the rates from film dosimetry. Differences are not too high, namely 1.9-2.4 times the combined standard deviation.
               