The effect of cooling rate on the precision of measurement of glow peak 5 in the thermoluminescence of LiF:Mg,Ti (TLD-100).

The effect of natural rapid cooling and oven slow cooling on the precision of thermoluminescence measurements of LiF:Mg,Ti is investigated. Three separate series of measurements resulted in average precisions of… Click to show full abstract

The effect of natural rapid cooling and oven slow cooling on the precision of thermoluminescence measurements of LiF:Mg,Ti is investigated. Three separate series of measurements resulted in average precisions of 5.1 and 5.0%, respectively. However, the highest precision of 1.7% (1 SD) was achieved for an oven-cooled material.

Keywords: thermoluminescence; effect; precision; rate precision; cooling rate; effect cooling