Calculation of the maximum allowed ambient dose rate outside CT rooms to quantitatively assess the structural shielding performance

Photo from wikipedia

The goal of this study was to develop a method to quantitatively assess the structural shielding performance of computed tomography (CT) rooms in Switzerland. The method was based on the… Click to show full abstract

The goal of this study was to develop a method to quantitatively assess the structural shielding performance of computed tomography (CT) rooms in Switzerland. The method was based on the comparison between calculated maximum allowed ambient dose rate (DR) and measured ambient DR in adjacent spaces. If the measured DR exceeds the calculated maximum allowed value, additional shielding is required; otherwise the shielding is sufficient. The maximum allowed ambient DR was calculated using two different shielding concepts either based on the tube loading of the scanner or on the accumulated dose length product of the scans. The method was validated for 36 spiral CT head protocols. The average ratio of the maximum allowed ambient DR between both concepts was 1.11 (range 0.57-2.14). Among 36 CT rooms, 7 room boundaries were insufficiently shielded. In conclusion, this method allows the authority to check on-site the compliance of the structural shielding with legal requirements.

Keywords: allowed ambient; assess structural; maximum allowed; structural shielding; quantitatively assess

Journal Title: Radiation Protection Dosimetry
Year Published: 2017

Link to full text (if available)


Share on Social Media:                               Sign Up to like & get
recommendations!

Related content

More Information              News              Social Media              Video              Recommended



                Click one of the above tabs to view related content.