WALL THICKNESS OPTIMISATION OF AN IONISATION CHAMBER FOR DIRECTIONAL DOSE EQUIVALENT RATE MEASUREMENT AT LOW AND MEDIUM PHOTON ENERGIES.

A thin and plane wall ionisation chamber having 900 cc volume was designed and fabricated to study the calibration coefficient dependency on ionisation chamber wall thickness for directional dose equivalent… Click to show full abstract

A thin and plane wall ionisation chamber having 900 cc volume was designed and fabricated to study the calibration coefficient dependency on ionisation chamber wall thickness for directional dose equivalent rate (H'(d)) at various low and medium energy X-ray beam qualities. Optimised wall thickness was established through measurements to achieve a near flat energy response using the developed ionisation chamber. The measurement shows that in the energy range 12-213 keV, the average calibration coefficient for directional dose equivalent rate lies within ±10% for wall thickness 1.2 mg/cm2 and 480 mg/cm2 (4 mm poly methyl methacrylate) for skin and eye lens dose monitoring, respectively. The study could be useful for the estimation of skin and eye lens doses for the paramedical staff and patient during the interventional radiology and interventional cardiology procedures by monitoring the directional dose equivalent rate.

Keywords: equivalent rate; ionisation chamber; dose equivalent; directional dose; wall thickness

Journal Title: Radiation protection dosimetry
Year Published: 2018

Link to full text (if available)


Share on Social Media:                               Sign Up to like & get
recommendations!

Related content

More Information              News              Social Media              Video              Recommended



                Click one of the above tabs to view related content.