In this work, the initial results of the type testing of the LiF:Mg,Cu,P (TLD-100H) whole-body personal dosemeters are presented. An assessment of reproducibility, linearity of the response, the residual signal… Click to show full abstract
In this work, the initial results of the type testing of the LiF:Mg,Cu,P (TLD-100H) whole-body personal dosemeters are presented. An assessment of reproducibility, linearity of the response, the residual signal as a function of the dose, energy and angular dependence of the response was performed. In general, the dosemeters show good reproducibility for different dose values and a linear behaviour for a range between 0.1 and 300 mSv. The detection limits obtained are lower than 50 μSv. The system presents a good energy and angular response for different radiation qualities.
               