Aerosol and droplet creation during oscillatory motion of the microkeratome amidst COVID-19 and other infectious diseases

Photo by brandsandpeople from unsplash

High-speed shadowgraphy showed droplets generated during flap cut with microkeratome. However, aerosolization of these droplets was unlikely because these droplets would settle on surface before aerosolization. Click to show full abstract

High-speed shadowgraphy showed droplets generated during flap cut with microkeratome. However, aerosolization of these droplets was unlikely because these droplets would settle on surface before aerosolization.

Keywords: microkeratome; oscillatory motion; aerosol droplet; motion microkeratome; creation oscillatory; droplet creation