LAUSR.org creates dashboard-style pages of related content for over 1.5 million academic articles. Sign Up to like articles & get recommendations!

Influence of noninvasive central blood pressure devices for afterload monitoring with aortic velocity-pressure Loop in anesthetized patients

Photo from wikipedia

Background Global afterload angle (GALA) is a parameter derived from velocity-pressure loop (VP Loop), for continuous assessment of cardiac afterload in the operating room. It has been validated with invasive… Click to show full abstract

Background Global afterload angle (GALA) is a parameter derived from velocity-pressure loop (VP Loop), for continuous assessment of cardiac afterload in the operating room. It has been validated with invasive measure of central pressure. The aim of this study was to evaluate the feasibility of noninvasive VP Loop obtained with central pressure measured with two different noninvasive tonometers. Methods A prospective, observational, monocentric study was conducted in 51 patients under general anesthesia. Invasive central pressure (cPINV) was measured with a fulfilled intravascular catheter, and noninvasive central pressure signals were obtained with two applanation tonometry devices: radial artery tonometry (cPSHYG: Sphygmocor tonometer) and carotid tonometry (cPCOMP: Complior tonometer). Three VP Loops were built: VP LoopINV, VP LoopSPHYG and VP LoopCOMP. Patients were separated according to cardiovascular risk factors. Results In the 51 patients under general anesthesia, cPSHYG was adequately obtained in 48 patients (89%) but, compared to cPINV, SBP was underestimated (−4 ± 6 mmHg, P < 0.0001), augmentation index (AIXSPHYG) and a GALASPHYG were overestimated (+13 ± 19%, P = 0.0077 and +4 ± 8°, P = 0.0024, respectively) with large limit of agreement (LOA) (−21 to 47% and −13 to 21° for AIXSPHYG and GALASPHYG, respectively). With the Complior, the failure rate of measurement for cPCOMP was 41%. SBP was similar (3 ± 17 mmHg, P = 0.32), AIXCOMP was underestimated (−11 ± 19%, P = 0.0046) and GALACOMP was similar but with large LOA (−50 to 26% and −20 to 18° for AIXCOMP and GALACOMP, respectively). Conclusion In anesthetized patient, the reliability of noninvasive central pressure monitoring by tonometry seems too limited to monitor cardiac afterload with VP Loop.

Keywords: pressure loop; blood pressure; central pressure; pressure; velocity pressure; noninvasive central

Journal Title: Blood Pressure Monitoring
Year Published: 2020

Link to full text (if available)


Share on Social Media:                               Sign Up to like & get
recommendations!

Related content

More Information              News              Social Media              Video              Recommended



                Click one of the above tabs to view related content.