LAUSR.org creates dashboard-style pages of related content for over 1.5 million academic articles. Sign Up to like articles & get recommendations!

Multicenter Validation of a Urine CXCL10 Assay for Noninvasive Monitoring of Renal Transplants.

BACKGROUND Urine CXCL10 (C-X-C motif chemokine ligand 10, interferon gamma-induced protein 10 [IP10]) outperforms standard-of-care monitoring for detecting subclinical and early clinical T-cell-mediated rejection (TCMR) and may advance TCMR therapy… Click to show full abstract

BACKGROUND Urine CXCL10 (C-X-C motif chemokine ligand 10, interferon gamma-induced protein 10 [IP10]) outperforms standard-of-care monitoring for detecting subclinical and early clinical T-cell-mediated rejection (TCMR) and may advance TCMR therapy development through biomarker-enriched trials. The goal was to perform an international multicenter validation of a CXCL10 bead-based immunoassay (Luminex) for transplant surveillance and compare with an electrochemiluminescence-based (Meso Scale Discovery [MSD]) assay used in transplant trials. METHODS Four laboratories participated in the Luminex assay development and evaluation. Urine CXCL10 was measured by Luminex and MSD in 2 independent adult kidney transplant trial cohorts (Basel and TMCT04). In an independent test and validation set, a linear mixed-effects model to predict (log10-transformed) MSD CXCL10 from Luminex CXCL10 was developed to determine the conversion between assays. Net reclassification was determined after mathematical conversion. RESULTS The Luminex assay was precise, with an intra- and interassay coefficient of variation 8.1% and 9.3%; showed modest agreement between 4 laboratories (R 0.96 to 0.99, P < 0.001); and correlated with known CXCL10 in a single- (n = 100 urines, R 0.94 to 0.98, P < 0.001) and multicenter cohort (n = 468 urines, R 0.92, P < 0.001) but the 2 assays were not equivalent by Passing-Bablok regression. Linear mixed-effects modeling demonstrated an intercept of -0.490 and coefficient of 1.028, showing Luminex CXCL10 are slightly higher than MSD CXCL10, but the agreement is close to 1.0. After conversion of the biopsy thresholds, the decision to biopsy would be changed for only 6% (5/85) patients showing acceptable reclassification. CONCLUSION These data demonstrate this urine CXCL10 Luminex immunoassay is robust, reproducible, and accurate, indicating it can be readily translated into clinical HLA laboratories for serial posttransplant surveillance.

Keywords: validation urine; multicenter validation; cxcl10 assay; cxcl10; urine cxcl10

Journal Title: Transplantation
Year Published: 2023

Link to full text (if available)


Share on Social Media:                               Sign Up to like & get
recommendations!

Related content

More Information              News              Social Media              Video              Recommended



                Click one of the above tabs to view related content.