Objective: Two-dimensional (2D) photoacoustic (PA) imaging based on array transducers provide high spatial resolution in the lateral direction by adopting receive dynamic focusing. However, the quality of PA image is often deteriorated by poor elevational resolution which is achieved by an acoustic lens. To overcome this limitation, we present a three-dimensional (3D) image reconstruction method using a commercial one-dimensional (1D) array transducer. Methods: In the method, the elevational resolution is improved by applying synthetic aperture focusing (SAF) technique along the elevational direction. For this, a commercially available 1D array transducer with an acoustic lens is modeled and appropriate synthetic focusing delay that can minimize the effect of the acoustic lens is derived by mathematical analysis. Results: From the simulation and experiment results, it was demonstrated that the proposed method can enhance the image quality of PA imaging, i.e., elevational resolution and signal-to-noise ratio (SNR). Conclusion: 3D PA images with improved elevational resolution were achieved using a clinical 1D array transducer. Significance: The presented method may be useful for clinical application such as detecting microcalcification, imaging of tumor vasculature and guidance of biopsy in real time.
               