X-ray sources have important applications in medical diagnosis, industrial inspection, etc. Transparent X-ray source has not been reported before, which enables the applications such as simultaneous imaging using X-ray and… Click to show full abstract
X-ray sources have important applications in medical diagnosis, industrial inspection, etc. Transparent X-ray source has not been reported before, which enables the applications such as simultaneous imaging using X-ray and visible light. In this article, a fully vacuum-encapsulated transparent flat-panel X-ray source was prepared by using ZnO nanowires cold cathode and indium-tin-oxide (ITO) transmission anode. The effect of thicknesses of ITO film on X-ray radiation intensity was studied. A radiation dose rate of 6.85 mGy/s measured at 5 cm in front of the flat-panel X-ray source (FPXS) under 48-kV anode voltage and an optical transmittance of 72.63% at 550-nm light were achieved. X-ray spectra show the indium is the main element to generate the X-ray. By adopting the device in an X-ray and visible light dual-functional inspection set-up, the X-ray and visible light imaging of objects were demonstrated simultaneously.
               