While the simultaneous scanning measurement and temperature field reconstruction of multiple aero-engine turbine components are extremely important for engine condition monitoring and parameter control, it is currently difficult to achieve.… Click to show full abstract
While the simultaneous scanning measurement and temperature field reconstruction of multiple aero-engine turbine components are extremely important for engine condition monitoring and parameter control, it is currently difficult to achieve. This study proposed a panoramic scanning optical pyrometer with circumferential and radial scan ranges of 0°–360° and 0–80 mm, respectively. Moreover, an innovative 2-D temperature field reconstruction algorithm based on panoramic scanning was proposed. After the temperature calibration of the pyrometer, the steady-state temperature measurement error was less than 1.6 °C, and the transient temperature measurement error was less than 2.5 °C. Furthermore, scanning measurement experiments were performed to verify the uncertainty of the 2-D temperature field reconstruction algorithm and the panoramic reconstruction function of the proposed pyrometer.
               