Balun or trap circuits are critical components for suppressing common-mode currents flowing on the outer conductors of coaxial cables in RF coil systems for Magnetic Resonance Imaging (MRI) and Magnetic… Click to show full abstract
Balun or trap circuits are critical components for suppressing common-mode currents flowing on the outer conductors of coaxial cables in RF coil systems for Magnetic Resonance Imaging (MRI) and Magnetic Resonance Spectroscopy (MRS). Common-mode currents affect coils’ tuning and matching, induce losses, pick up extra noise from the surrounding environment, lead to undesired cross-talk, and cause safety concerns in animal and human imaging. First proposed for microwave antenna applications, the Lattice balun has been widely used in MRI coils. It has a small footprint and can be easily integrated with coil tuning/matching circuits. However, the Lattice balun is typically a single-tuned circuit and cannot be used for multi-nuclear MRI and MRS with two RF frequencies. This work describes a dual-tuned Lattice balun design that is suitable for multi-nuclear MRI/MRS. It was first analyzed theoretically to derive component values. RF circuit simulations were then performed to validate the theoretical analysis and provide guidance for practical construction. Based on the simulation results, a dual-tuned balun circuit was built for 7T 1H/23Na MRI and bench tested. The fabricated dual-tuned balun exhibits superior performance at the Larmor frequencies of both 1H and 23Na, with less than 0.15 dB insertion loss and better than 17 dB common-mode rejection ratio at both frequencies.
               
Click one of the above tabs to view related content.